Lumax Technical Manual 33
Parameter
Range
Standard
AV-hysteresis mode
OFF,
positive,
negative
OFF
AV hysteresis
10..(10)..150 ms
[50 ms]
AV repetitive
(positive)
OFF; 1..(1)..10
[5]
AV repetitive
(negative)
OFF;
1..(1)..15..(5)..100..(10)..1
80
[180]
AV scan
OFF, 1..(1)..10
[5]
I-Opt
I-Opt
OFF, ON
OFF
AV hysteresis at I-
Opt
400 ms
400 ms
AV repetitive at I-Opt OFF; 1..(1)..10
[5]
AV scan at I-Opt
OFF, 1..(1)..10
[5]
AV max at I-Opt
400 ms
400ms
Post-ventricular Atrial Refractory Period (PVARP)
PVARP
175..(25)..600 ms, Auto
250 ms
Auto PVARP
OFF, ON
OFF
VES Classification (VES Lock-in Protection)
VES discrimination
after As
250..(50)..450 ms
350 ms
Rate Adaptation (Acceleration Sensor)
Maximum sensor
rate
90..(5)..160ppm
120ppm
Sensor gain
1.0; 1.1; 1.3; 1.4; 1.6; 1.8;
2.0; 2.2; 2.6; 3.0; 3.3; 3.7;
4.0; 4.5; 5.0; 6.0; 7.0; 8.0;
9.0; 10; 11; 12; 14; 16; 18;
20; 22; 24; 28; 32; 35; 40
6.0
Auto Sensor gain
OFF, ON
OFF
Sensor threshold
(numbers are not
displayed on screen)
Very low = 0
Low = 3
Mean = 7
High = 11
Very high = 15
Mean
Rate increase
0.5; 1..(1)..6 ppm/cycle
2 ppm/cycle
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......