Lumax Technical Manual 21
Complications and observations for the Tupos LV/ATx study
group are summarized in
Table 3
and
Table 4
. The total number
of patients may not equal the sum of the number of patients listed
in each category, as an individual patient may have experienced
more than one complication or observation.
Table 3: Summary of Complications – Tupos LV/ATx
Category
Number
of
Patients
% of
Patients
Number of
Complications
Complications
per patient-
year
Procedure Related
Hematoma
4
3.01%
4
0.04
Pneumothorax
2
1.50%
2
0.02
Total
6
4.51%
6
0.06
Atrial Lead Related
Dislodgement
3
2.26%
3
0.03
Total
3
2.26%
3
0.03
ICD Lead Related
High threshold/ No
capture
2
1.50%
2
0.02
Diaphragmatic/
Intercostal
stimulation (RV)
1
0.75%
1
0.01
Total
3
2.26%
3
0.03
LV Lead Related
High threshold/
Intermittent
biventricular capture/
No capture
11
8.27%
12
0.12
Unable to implant
lead via coronary
sinus
11
8.27%
11
0.11
Dislodgement
4
3.01%
4
0.04
Diaphragmatic/
Intercostal
stimulation
1
0.75%
2
0.02
Total
27
20.3%
29
0.28
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......