110 Lumax Technical Manual
2.11.2.4 Types of Report Transmissions
When the Home Monitoring function is activated, the transmission
of information from the implant can be triggered as follows:
•
Daily report – the time period (daily) initiates the
transmission (IEGM included if not yet sent)
•
Event report – the ICD/CRT-D detects certain events,
which initiate a transmission (IEGM included if not yet
sent)
•
Periodic IEGM report – certain event reports can be
programmed to have an IEGM included each time that
they are transmitted. The IEGM includes comprehensive
event details with up to 10 seconds of IEGM (8 seconds
pre-detection and 2 seconds pre-termination).
•
Programmer triggered report – a test message
transmitted upon request of the physician in the clinic.
Daily Report
The time of daily Home Monitoring Report transmission is
programmable. For periodic messages, the time can be set
anywhere between 0:00 and 23:59 hours. It is recommended to
select a time between 0:00 and 4:00.
Periodic Report with IEGM
The Lumax ICDs and CRT-Ds can be programmed to transmit
IEGMs with the daily Home Monitoring report on a periodic basis
with the interval selected; OFF, 2, 3, 4, or 6 months.
Event Report
When certain cardiac and technical events occur, a report is
automatically generated. This information is described as an
“event report.”
The implant supports the following automatic event triggers:
•
Termination of VT/VF Episode (not Termination of a
Monitoring episode, SVT episode or Detection while
Magnet applied (DTM) episode)
•
Ongoing Atrial Monitoring Episode lasting longer than the
programmed time (0.5, 6, 12 or 18 hours)
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......