
26 Lumax Technical Manual
Table 29 Lumax 300 HF-T Longevity Estimates
Shocks per year
100% Pacing Support
(in years)
4
5.4
5
5.3
6
5.2
7
5.1
8
4.9
9
4.8
10
4.7
11
4.6
12
4.5
Table 30 Lumax 340 HF-T Longevity Estimates
Shocks per year
100% Pacing Support
(in years)
4
5.2
5
5.0
6
4.8
7
4.7
8
4.6
9
4.4
10
4.3
11
4.2
12
4.1
The Lumax 300/340 HF CRT-Ds (Lumax
300
HF-T, and
Lumax 340 HF-T) are intended to operate for more than 5 years
under normal use.
5.2.2 Lumax 500/540 Devices
The single chamber Lumax 500/540 variants (Lumax 500 VR-T,
and Lumax 540 VR-T) of ICDs are intended to operate for more
than 6 years under normal use.
Table 31
provides longevity
estimates for these devices. The table provides several different
support scenarios. It is assumed that the shocks are equally
spaced throughout the life of the ICD.
Summary of Contents for Lumax DR ICD Series
Page 8: ...vi Lumax Technical Manual...
Page 60: ...52 Lumax Technical Manual...
Page 135: ...6 Lumax Technical Manual...
Page 143: ...14 Lumax Technical Manual Figure 11 Header Ports...
Page 151: ...22 Lumax Technical Manual...
Page 168: ...Lumax Technical Manual 39...
Page 170: ...Lumax Technical Manual 41...
Page 175: ......