36 Lumax Technical Manual
Parameter
Range
Standard
Ventricular Therapy Parameters
Energy 1st shock
and 2nd shock
VT1, VT2 (model
300/500)
1
OFF; 1..(1)..16..(2)..30 J
30 J
Energy 1st shock
and 2nd shock
VF (model 300/500)
1
1..(1)..16..(2)..30 J
30 J
Energy 1st shock
and 2nd shock
VT1, VT2 (model
340/540)
1
OFF; 1..(1)..16..(2)..40 J
40 J
Energy 1st shock
and 2nd shock
VF (model 340/540)
1
1..(1)..16..(2)..40 J
40 J
Number of shocks
(VT1/VT2)
0..(1)..8
[8]
Number of shocks
(VF)
2; 6..(1)..8
8
Confirmation (per
zone)
OFF, ON
ON
Shock form (per
zone)
Biphasic, biphasic2
Biphasic
Polarity (per zone)
Normal,
Reverse,
Alternating
Normal
ATP Parameters
ATP type
Burst, Ramp, Burst + PES
[Burst]
ATP attempts
OFF, 1…(1)...10
3
S1 number
1...(1)...10
[5]
Add. S1
OFF, ON
[ON]
R1-S1 interval
200...(10)...500 ms
(absolute);
70...(5)...95 % (adaptive)
[80 %]
S1 (RAMP)
decrement
5...(5)...40 ms
[10 ms]
Scan decrement
OFF, 5...(5)...40 ms
[OFF]
1
Programmed values are stored energies.