Rev. 5.01
130
been on. (LoFlo only)
A-4-8.
Pump max time
This value indicates the maximum
rated lifetime of the sampling pump.
(LoFlo only)
Min. display
A-5.
Module setup
Module setup menu
A-5-1.
Current period
This setting is used to set the
calculation period of the ETCO2 value.
The end-tidal CO2 value is the highest
peak CO2 value of all end of
expirations (end of breaths) over the
selected time period. If less than two
breaths exist in the selected time
period, the value will be the maximum
ETCO2 value for the last two breaths.
1 breath,
10 sec,
20 sec
A-5-2.
Balance gas
setup the gas in the measurement;
the type of gas that is mixed with the
breathing gas measuring
Room air
N2O
Helium
A-5-3.
Sleep mode
Sleep mode is used to save power
when the host monitor is in standby
mode. There are two sleep modes
available for the Capnostat. Using
Sleep Mode 1 maintains the heaters so
the Capnostat is able to run
immediately after exiting the sleep
mode. Mode 2 will require the
Capnostat to go through its warmup
sequence when exiting this mode and
a delay will be introduced until the
system has stabilized.
Normal mode
Turnoff mode
Power saving
A-5-4.
BARO. pressure
This setting is used to set current
Barometric Pressure.
760mmHg
A-5-5.
Gas temperature
This setting is used to set temperature
of the gas mixture. This setting is
useful when bench testing using static
35.0
°
C
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...