Rev. 5.01
107
condition
SpO2 - checkprobe
SpO2 - poorsignal
SpO2 - lostpulse
SpO2 - artifact
SpO2 - pulse search
A-2.
Rate volume
Menu in which rate volume is set up
when the ECG volume is set, it is
automatically set to Off.
Off, 0%~100%
A-3.
Display option
SpO2 waveform display setting
A-3-1.
Sweep speed
It can set the speed of SpO2 displayed
on the screen. Default: 25 mm/s.
6.25mm/s, 12.5mm/s,
25mm/s, 50mm/s
SpO2 Status Messages
Below is a list of system status alarm messages which may be displayed in the SPO2 parameter
window during monitoring.
Message
State
Check probe
Occurs when a probe separates from an animal. Check out the
Probe.
Pulse search
Detection by the monitor of a repeatable pulse has ceased. Check
the animal and the probe site.
Poor signal
The SpO2 signal is too low. No SPO2 data is displayed. This can
be due to a low veterinary animal pulse, animal motion, or some
other interference. Check the animal and the probe.
Lost signal
SpO2 data continues to be displayed, but the quality of the signal
is questionable. Check the animal and the probe.
Artifact
It indicates that something happened to the pulses; determine if
the artifact to be abnormal and irregular
Cleaning
⚫
Do not autoclave, pressure sterilizes, or gas sterilizes.
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...