Rev. 5.01
172
M
anufacturer’s
Declaration - Electromagnetic Emission
The BM3Vet Elite, BM5Vet Elite, BM7Vet Elite system are intended for use in the electromagnetic
environment specified below. The customer or the user of BM3Vet Elite, BM5Vet Elite, BM7Vet Elite
system should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
Mains terminal disturbance
voltage CISPR 11
GROUP1, CLASS A
The EMISSIONS characteristics of BM3Vet Elite,
BM5Vet Elite, BM7Vet Elite make it suitable for use
in industrial areas and hospitals (CISPR 11 class A). If
it is used in a residential environment (for which
CISPR 11 class B is normally required) BM3Vet Elite,
BM5Vet Elite, BM7Vet Elite might not offer adequate
protection to radio-frequency communication
services. The user might need to take mitigation
measures, such as relocating or re-orienting the
equipment.
RADIATED DISTURBANCE C
ISPR 11
GROUP1, CLASS A
Harmonic Current Emission
IEC 61000-3-2
CLASS A
The BM3Vet Elite, BM5Vet Elite, BM7Vet Elite are
suitable for use in all establishments other than
domestic and may be used in domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Voltage change, Voltage
fluctuations and Flicker
Emission IEC 61000-3-3
Pst: 1
Plt: 0.65
Tmax: 0.5
dmax: 4%
DC: 3.3%
M
anufacturer’s Declaration
- Electromagnetic Immunity
The BM3Vet Elite, BM5Vet Elite, BM7Vet Elite system are intended for use in the electromagnetic
environment specified below. The customer or the user of the BM3Vet Elite, BM5Vet Elite, BM7Vet
Elite system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -
guidance
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...