A-1-1.
Arrhythmia type
Arrhythmia diagnosis level setup menu.
OFF: Do not perform arrhythmia
diagnosis.
LETHAL: Performs the detection of
Asys, VTAC/VFIB, and VTAC at the
selected lead
FULL: Performs the detection of all 13
arrhythmia. (Only for BM5Vet
Elite/BM7Vet Elite)
OFF, LETHAL, FULL
A-1-2.
Arrhythmia alarm
condition
Alarm setting menu by arrhythmia type
Warning
VENTRICULAR ARRHYTHMIAS
The arrhythmia analysis program is intended to detect ventricular
arrhythmia. This program is not designed to detect trial or supra ventricular
arrhythmias. In some cases, it may not be possible to distinguish the
presence or absence of arrhythmias. Therefore, doctors should analyze the
arrhythmia information like other medical information.
SUSPENDED ANALYSIS
Certain conditions can delay the arrhythmia analysis. Detection and alarms
associated with arrhythmias do not occur when arrhythmia conditions are
delayed. This message is generated when the arrhythmia analysis is delayed:
LEADS FAULT, ALARM PAUSE, ALL ALARMS OFF, DISCHARGED.
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...