Rev. 5.01
146
⚫
Enflurane
⚫
Halothane
⚫
Desflurane
Feature
Specification
CO2
Range
0
–
100 mmHg; 0
–
13.3 kPa; 0-10% CO2 STPD (standard
temperature and pressure dry)
Accuracy
± (0.2 vol% + 4% relative)
Rise Time
400 ms (average)
Anesthetic Agents
Gases
Isoflurane, Enflurane, Halothane, Desflurane, Sevoflurane
Range
Iso., Enf., Hal., Sev.: 0
–
6%
Des.: 0
–
18%
Resolution
0 .01%
Accuracy
± (0.15% vol% + 4% relative)
Rise Time
450 ms (average)
Calibration
Factory calibrated
Power Up Time
30 sec
Delay Time
< 4 sec
Dimensions
175mm x 85mm x 50mm (L x W x H)
Respiration
Range
0-150 breaths/min
Accuracy
±1 breath/min
Flow Rate
Range
170 ml/min
Accuracy
±20ml/min
Environment Condition
Operating Temperature
15
–
35°C
Storage Temperature
-
5 to 50°C
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...