Check for proper cuff size.
2.
Check for residual air left in the cuff from a previous measurement.
3.
Make sure cuff is not too tight or too loose.
4.
Make sure cuff and heart are at same level, otherwise hydrostatic
pressure will offset the NIBP value.
5.
minimize patient movement during measurement.
6.
Check for leak in cuff or tubing.
7.
Patient may have a weak pulse.
Measurement Limitations
The measurement may be inaccurate or impossible:
⚫
Measurements are impossible with heart rate extremes of less than 40 bpm or greater than
300 bpm, or if the patient is on a heart-lung machine.
⚫
With excessive and continuous patient movement such as shivering or convulsions
⚫
if a regular arterial pressure pulse is hard to detect
⚫
With cardiac arrhythmias
⚫
With rapid blood pressure changes
⚫
With severe shock or hypothermia that reduces blood flow to the peripheries
⚫
With obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming
from the artery
⚫
On an edematous extremity.
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...