Rev. 5.01
179
Respiration Performance
Method
Thoracic impedance
Channel Selection
RA-LL
BM5Vet Elite, BM7Vet Elite only: RA-LA
Measurement Range
5
–
120 breaths per minute
Accuracy
±
1 breath per minute
Apnea Alarm
Yes
SpO2 Performance
SpO2 Range
0 to 100%
SpO2 Accuracy
70 to 100%
±
2 digits
0 to 69% unspecified
Pulse Rate Range
18 to 450 bpm
Pulse Rate Accuracy
±
2 bpm
NIBP Performance
Method
Oscillometry with step deflation
Operation Mode
Manual/Automatic/Continuous
Measurement Range
Systolic: 40 to 260 mmHg
MAP: 26 to 220 mmHg
Diastolic: 20 to 200 mmHg
Accuracy
Mean error: less than
±
5 mmHg
Standard deviation: less than 8 mmHg