hardware error may exist.
CO2 OUT OF
RANGE
The value being calculated is greater
than the upper CO2
limit( 150mmHg )
If error persists, perform a zero.
CHECK AIRWAY
ADAPTER
Usually caused when the airway
adapter is removed from the
Capnostat or when there is an
optical blockage on the windows of
the airway adapter. May also be
caused by failure to perform
Capnostat zero to when adapter
type is changed.
To clear, clean airway adapter if
mucus or moisture is seen. If the
adapter is clean, perform a
Capnostat zero.
MODULE OFF
It occurs when the equipment and
module are separated. Message
output
Verify module connections
Service request
CO2 measurement failure
CO2 value is not output, or numerical error.
Troubleshoot procedure
1.
Check the connection between the main unit and the module.
2.
Check the module line connection with the filter line or airway.
3.
Replace filter line or airway.
4.
Service Request
Note
In the following monitoring conditions, the measured values may be
inaccurate. Read the measured values carefully.
1.
When using this in an environment of using nitrous oxide gas of high
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...