Rev. 5.01
141
IBP Settings
Menu
Description
Available settings
A.
IBP menu
A-1.
Alarm
IBP alarm settings menu
A-1-1.
Parameter alarm limit
IBP-SYS, MEAN, DIA, PR parameter
alarm, level, activate setup
A-1-2.
Technical
alarm
condition
IBP-cableoff
IBP-disconnect
A-2.
BP filter
Menu to set the filter to be applied
when measuring
OFF: 0Hz ~ 40Hz
12Hz: 0Hz ~ 12Hz:
Generally recommended for
monitoring
20Hz: 0Hz ~ 20Hz:
It is used to process higher frequency
waveform components, and when this
filter is used, the pressure value can
rise.
OFF
12Hz
20Hz
A-3.
Display option
IBP waveform display setting
A-3-1.
Sweep speed
Set the speed of IBP displayed on the
screen. The basic speed of the IBP is
25 mm/s.
6.25mm/s,
12.5mm/s,
25mm/s,
50mm/s
A-3-2.
IBP label
Menu to set measuring position
ART, FEM, PAP, RAP, LAP, UAP, UVP,
CVP, ICP, USERDEFINED
A-3-3.
Scale
Menu to set size of measurement
waveform on screen.
30, 60, 80, 100, 160,
200, 300 (in mmHg)
A-3-4.
Pulse rate
Menu to set display of blood pressure
pulse
B-1.
Zeroing
Menu to set zero-point of transducer.
If zero adjustment is not performed,
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...