Rev. 5.01
166
Never immerse the electrical connector in water or other liquids. When
cleaning, be careful not to let the liquid stick to the edge of the screen.
P
atient’s Cable
⚫
Clean the patient cables with a gauze pad moistened with a soap solution.
⚫
To disinfect patient cables, wipe the cables with a gauze moistened with diluted alcohol or
a glutaraldehyde-based dis-infectant.
⚫
Ethylene oxide is suitable for intensive disinfection (almost sterilization), but it shows that
the service life of cables and lead wires is reduced.
⚫
Dry thoroughly with a lint-free cloth.
Caution
Do not use disinfectants that contain phenol as they can spot plastics. Do not
autoclave or clean accessories with strong aromatic, chlorinated, ketone, ether,
or ester solvents. Never immerse electrical connectors.
When cleaning, do not apply excessive pressure or bend the cable
unnecessarily. Excessive pressure can damage the cable.
Reusable ECG Electrodes
Clean the electrode cup regularly with a toothbrush. When removing gel-like residues, use a soft
brush with flowing water. Wipe the electrode with a soapy cloth moistened with soapy water.
⚫
Sterilize the electrode by soaking the diluted alcohol in cloth.
⚫
Dry thoroughly with a lint-free cloth.
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...