Rev. 5.01
142
"
IMBALANCE
" message is displayed in
the waveform window.
Trouble Shooting
Trouble shootings for a case that blood pressure value is not displayed on screen
Description
Action to Take
In case of ‘
out of measurement range
’
situation
Check the measurement conditions.
In case blood pressure transducer is
damaged
Replace the damaged transducer with new one
Trouble shootings for a case the measured value is different from the expected value
Description
Action to Take
In case there are air bubbles in tubes
Remove the air bubbles
In case an extension tube is connected
Remove the extension tube
In case of using blood pressure transducer
with a different sensitivity
Check position of transducer
For other cases
Perform zero-point adjustment
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...