Rev. 5.01
173
Electrostatic Discharge
Immunity (ESD) IEC 61000-
4-2
±
8 kV/Contact
±2, ±4, ±8, ±15 kV/Air
±
8 kV/Contact
±2, ±4, ±8,
±15 kV/Air
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least 30%.
Radiated RF
Electromagnetic Field
Immunity IEC 61000-4-3
3 V/m 80 MHz - 2.7
GHz 80% AM at 1 kHz
3 V/m 80 MHz - 2.7
GHz 80% AM at 1 kHz
BM3Vet Elite, BM5Vet
Elite, BM7Vet Elite are
suitable to use in
professional healthcare
environment.
RF communication
equipment is used no
closer than 30 cm to
any part of the BM3Vet
Elite, BM5Vet Elite,
BM7Vet Elite, including
cables specified by
Bionet
Immunity to Proximity
Fields from RF wireless
Communication s
Equipment IEC 61000-4-3
28 V/m Max. 3855785
MHz in according to
table 9 in IEC 606011-2
28 V/m Max. 3855785
MHz in according to
table 9 in IEC 606011-2
Mains power quality
should be that of a
typical commercial or
hospital environment.
Electrical Fast
Transient/Burst Immunity
IEC 61000-4-4
±
2 kV, 100 kHz
repetition frequency
±
2 kV, 100 kHz
repetition frequency
The quality of supplied
power should be
suitable for general
business site or
hospital environment.
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...