Rev. 5.01
114
is measured, pressurize to the set
pressure value every time
A-5.
Measurement Interval
A menu to set Interval time when the
blood pressure is measured.
After setting INTERVAL, you must press
NIBP KEY to start NIBP measurement.
1min, 2, 3, 4, 5, 10,
15, 20, 30, 60, 90,
2 hours, 4, 8
A-6.
Vital sign review
Record the last 40 blood pressure
readings.
Warning
Check periodically to see if the circulation from the cuff to the distal part of
the
animal’s
limb is good.
1minute and 2minute intervals when using automatic measurement, check the
animal’s
condition frequently. It is not recommended for measuring blood
pressure for a long time if the measurement time period is set to 10minutes
or less.
When changing the cuff, be sure to adjust the actual size of the cuff and the
size of the cuff set on the equipment.
Note
Safety Considerations
Software and Hardware for Cuff pressure Blocking
The cuff is automatically deflated when the measurement time is longer than
two minutes in Cat/ Dog/ Horse mode and more than 90seconds in Puppy
mode. Extension limits are set for all animal categories to prevent overpressure
on the animal.
The maintenance is performed every 2 years
(See maintenance recommendations section in this manual)
Check the following list of devices to always operate properly and safety.
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...