Rev. 5.01
105
Note
SpO2 wave size changes automatically
Signal and Data Validity
It is extremely important to determine that the probe is attached to the animal correctly and the
data is verifiable. To make this determination, three indications from the monitor are of assistance
—
signal strength bar, quality of the SPO2 waveform, and the stability of the SPO2 values. It is critical
to observe all three indications simultaneously when ascertaining signal and data validity.
Signal Strength Bar
The signal strength bar is displayed within the SPO2 values window. This bar consists of 10 blocks
set depending on the strength of the signal. Proper environmental conditions and probe attachment
will help to ensure a strong signal.
Quality of SPO2 Waveform
Under normal conditions, the SPO2 waveform corresponds to (but is not proportional to) the arterial
pressure waveform. The typical SPO2 waveform indicates not only a good waveform, but helps the
user find a probe placement with the least noise spikes present. The figure below represents an
SPO2 waveform of good quality.
Good Quality SPO2 Waveform
If noise (artifact) is seen on the waveform because of poor probe placement, the photo detector
may not be flush with the tissue. Check that the probe is secured and the tissue sample is not too
thick. Pulse rate is determined from the SPO
2
waveform which can be disrupted by a cough or other
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...