Rev. 5.01
92
A-5-6.
Initial setup
ISO, ST point position initial value
setting Menu
ISO: 80
ST: 108
A-6.
Pacemaker
Pacemaker detection display setting
ON/OFF
Trouble Shooting
Problem
Solution
Inaccurate heart rate
and/or false asystole.
Check ECG signal from patient:
1.
Check/adjust lead placement.
2.
Check/perform skin preparation.
3.
Check/replace electrodes.
Check amplitude of ECG waveform:
1.
Select ECG parameter label.
2.
Select DISPLAY LEAD,
3.
Scroll through all ECG leads and check for 0.5mV
amplitude at normal (1X) size. (at least 0.5mV amplitude is
required for QRS detection.) for borderline signals, validate
on a graph.
4.
If amplitudes are low, electrodes may need to be
repositioned or replaced.
False ventricular fibrillation
occurrence
Check ECG signal from patient: (the chest lead may exhibit polarity
changes which may occasionally cause an inaccurate call.)
1.
Check/adjust electrode placement.
2.
Check the electrode attachment to the skin and attach it
correctly.
3.
Check the condition of the electrode and replace it if
necessary. (if chest lead is a problem, move the chest
electrode to another chest position or leg position.)
Inaccurate pacemaker
detection
Use pacemaker processing:
1.
Select ECG parameter label.
2.
Display the lead of ECG with the greatest amplitude in the
Summary of Contents for BM Elite Series
Page 21: ...Rev 5 01 21 Left side view Printer ...
Page 25: ...Rev 5 01 25 Left side view Printer ...
Page 29: ...Rev 5 01 29 Left side view Printer USB port ...
Page 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Page 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Page 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Page 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Page 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Page 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Page 161: ...Rev 5 01 161 Power failure Print failure ...
Page 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Page 163: ...Rev 5 01 163 Periodic Noises ...
Page 215: ...Rev 5 01 215 BIONET CO LTD ...