Section 12 Radiation Protection
111
□
Unscrew screw from old source and close transport
shielding.
□
Open transport shielding of new source.
□
Turn screw into new source.
□
Take hold of rod source using a pair of pliers and insert it
into the shielding until it touches the other end.
□
Turn screw out of new source.
□
Screw down cover cap using a large screwdriver.
□
Attach cover plate.
□
Insert emergency rod again.
□
Lock with padlock.
□
Check proper OPEN/CLOSED function of locking mecha-
nism of radiation exit channel (other side of shielding).
□
Replace identity plate on shielding or attach new source
number.
□
Observe the special regulations regarding labeling and
transport of the shielding containing the old source back
to the manufacturer.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...