Section 9 Operation of the Radiometric
Bulk Flow Meter
79
Minute printout
Minute: 13/08/97 13:16 P1
Weight[t]
3.617 Diff[t] 0.032
Minute: 13/08/97 13:17 P1
Weight[t]
3.661 Diff[t] 0.022
Minute: 13/08/97 13:18 P1
Weight[t]
3.680 Diff[t] 0.013
Minute: 13/08/97 13:19 P1
Weight[t]
3.699 Diff[t] 0.029
Minute: 13/08/97 13:20 P1
Weight[t]
3.725 Diff[t] 0.031
Minute: 13/08/97 13:21 P1
Weight[t]
3.710 Diff[t] 0.024
etc.
Shift printout
Hour: 13/08/97
07:00
P1
Weight[t]
3793.617
Total
[t]
3793.617
Hour: 13/08/97
08:00
P1
Weight[t]
3855.234
Total
[t]
7648.851
Hour: 13/08/97
09:00
P1
Weight[t]
3639.364
Total
[t]
11288.215
Hour: 13/08/97
10:00
P1
Weight[t]
3868.222
Total
[t]
15156.437
Hour: 13/08/97
11:00
P1
Weight[t]
3697.430
Total
[t]
18853.867
Hour: 13/08/97
12:00
P1
Weight[t]
3894.555
Total
[t]
22748.422
Hour: 13/08/97
13:00
P1
Weight[t]
3633.837
Total
[t]
26382.259
Hour: 13/08/97
14:00
P1
Weight[t]
3793.943
Total
[t]
30176.202
Shift: 13/08/97
14:00
P1
Weight[t]
30176.202Total
[t]
30176.202
etc.
9.2.2 Zero Point Correction through Tare Measurements
In addition to the on-going measurement which is displayed live,
the tare measurement is also part of normal operation.
All changes occurring with time must be compensated for by a tare
measurement. In particularly, this includes changes on the belt it-
self. A deposit layer may also form on the belt depending on the
material being conveyed. The user must decide how often the tare
weight should be measured. The button must only be pressed
when the belt is absolutely empty, everything else is done auto-
matically. We recommend checking the tare once a month.
Tare must also be determined again when the belt has been re-
placed. Recalibration is not necessary.
The same applies if the source is replaced after years (see section
Radiation Protection Guidelines when this should be done).
No tare measurement is necessary to compensate for the natural
decay of activity. Since the computer also contains a clock (
Date &
Time
in the
General Data
menu), this change can be compensated
for mathematically. It takes place every night exactly at midnight.
You can check this as well. After several days the tare count rate
changes automatically. If the computer has been without power
supply for some time, the clock is supplied with power by a bat-
tery. At the next measurement, the decay compensation is cor-
rected automatically.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...