9.3 Operation with Different Materials
80
9.2.3 Batch Operation
In addition to weighing the conveyed material continuously, the
bulk flow measurement can also be used for batch weighing. This
operation mode allows, for example, to load trucks with a given
amount, or to control a feeding machine.
The following settings are required:
o
In the
Measuring Parameter
menu, select the product being
measured.
o
In the
Measuring Parameter
menu, select as
Operating Mode
the
mode
Batch via keyboard
or
Batch via external
.
o
In the I/O Config. menu, select
Clear Ext.
Batch Start/Mass In-
tegrator for Digital Input 1
if you want to perform an external
batch measurement.
o
In the I/O Config. menu, select
Batch Ready
for
Digital Output
relay 3
and then enter the batch weight.
If the preset batch has passed through the measuring frame, a
signal is triggered at digital output relay 3 that can be used, for
example, to turn off the belt.
Another batch measurement can be carried out.
9.3 Operation with Different Materials
If various products requiring different calibrations are to be con-
veyed, you have to run separate calibrations for each product
(max. 10 different products).
The product can be selected either via keyboard by entering the
product number in the
Measuring Parameter
menu, item
Measuring
Product.
External selection is possible only for 2 different products
via digital input 3 (open = product 1, closed = product 2).
In the print protocol, the products are identified by the labels P1,
P2 ... P10.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...