Section 9 Operation of the Radiometric
Bulk Flow Meter
83
9.6.3 Error Messages during Calibration
Error
No. Error text
Cause
Solution
10
Not calibrated
No calibration coefficient
available or not calculated
- Enter coefficient or calculated it
38
Wrong product
If the product is unequal to
the calibration product when
reading-in the calibration
count rate
- Adapt product to calibration
product
41
min. measuring range not
reached
Enter negative measuring
range
- Check final value
(only with square or cubic curve)
42
max. measuring range
exceeded
Measuring range input lies
above the limit
- Check final value
(only with square or cubic curve)
43
Rate Overflow
With twisted measuring
range
- Check input
47
Check external mass
counter setting
Input is outside the limits
and is ignored.
(triggered when changing
parameters – measuring
range or mass counter)
- Increase step-width of mass
counter or reduce maximum
measuring range
(safety circuit, so relay output
can still follow)
9.6.4 Error Messages during Measurement
Error
No. Error text
Cause
Solution
1
Rate Overflow
Count rate
>520.000cps
- Check configuration of applica-
tion, or calculation
2
No Detector Rate
Error in detector
- Error in detector
- Multiplier or probe electronics
faulty
3
Wrong HV
HV < 500V or >1500V, out-
side the control range
- Error in detector
- Multiplier or probe electronics
faulty
6
Detector temperature >
65°C
Detector temperature has
exceeded 65°C
- Detector too warm, provide
adequate cooling
7
Trouble with detector
communication
Data transfer between de-
tector and evaluation unit
faulty
- Check connection terminals of
cable
- Detector or evaluation unit
faulty
- Broken cable
9
Detector temperature >
85°C
Detector temperature has
exceeded 85°C
- Detector too hot, provide ade-
quate cooling
44
Current input error
Input current above 23 mA - Check power source
46
Check external mass
counter setting
Input is outside the limits
and is ignored.
(triggered when changing
parameters – measuring
range or mass counter)
- Increase step-width of mass
counter or reduce maximum
measuring range
(safety circuit, so relay output
can still follow)
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...