Section 6 Evaluation Unit LB 442
53
Time Constant 1 1 (0 to 9999,9 sec):
Enter the time constant in seconds. The computer calculates a slid-
ing average of the area weight over this time. The higher the time
constant entered, the lower the statistical variations of the output
signal, the longer, however, the response to sudden changes
-
Time constant 2
(Only available via the button combination
SK1 & clear
.)
This time constant smoothes the statistics of the count rate prior to
the calculation and is independent of time constant 1. In most
cases, the standard value of 1 s is correct. Any change of this time
constant should be made only after consultation with BERTHOLD
TECHNOLOGIES.
-
Time constant for tacho or moisture input
(Available via
Time constant 2 + <more>
.)
This time constant smoothes the current input. In most cases, the
standard value of 1 s is correct. A higher value will smooth the
measured value at the current input more, but at the same time
delays the response to the load.
Curr. Input
(Only when operating mode
with moisture compensation
has
been selected)
Select the initial value 0 or 4 mA of the current input signal
(28a/28c).
Curr. Input limits
(Only when operating mode
with moisture compensation
has
been selected)
0/4mA
lower limit value in % moisture
20 mA
upper limit value in % moisture
or
(Only when operating mode
with tachometer with current in-
put
has been selected)
0/4mA
lower limit value in m/s
20 mA
upper limit value in m/s
Belt speed (0 to 99,99 m/s)
(Only when operating mode
constant belt speed
or
with moisture compensation
has been selected)
Enter the constant belt speed in m/s.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...