Section 11 Service Instructions
93
11.4.2 Error Code Listing
No.
Error text
Cause
Solution
1
Pulse rate overflow Pulse rate > 520,000 cps
- Check configuration of application, or
calculation
2
No pulses from
detector
Fault in detector
Error in detector
-Multiplier or probe electronics faulty
3
HV wrong
HV < 500V or >1500V, outside
controlled range
- Error in detector
-Multiplier or probe electronics faulty
5
Measuring range
for current output
0/4-20mA faulty
Upper limit is smaller or equal to
lower limit (triggered during cali-
bration calculation)
- Check measuring range input
6
Detector tempera-
ture > 65°C
Detector temperature has ex-
ceeded 65°C
- Detector too warm, provide adequate
cooling
7
Trouble with detec-
tor communication
Data transfer between detector
and evaluation unit faulty
- Check connection terminals of cable
-Detector or evaluation unit faulty
-Broken cable
9
Detector tempera-
ture > 85°C
Detector temperature has ex-
ceeded 85°C
- Detector too hot, provide adequate
cooling
10
No calibration
No calibration coefficient available
or not calculated
- Enter or calculate coefficient
11
Power fail
>1 month – time?
The date has changed since the
last operation by one month
- Check date
-Battery still OK?
12
No input allowed
Instrument locked by password
- Enter password
32
Wrong password
Wrong password input
- Enter correct password
34
Wrong input value Wrong parameter entry
- Input lies outside range
35
Calibration error
Parameters for calculation of cali-
bration curve are missing
- Check absorption coefficient and tare
38
Wrong product
Measurement product is not the
same as product while reading the
calibration count rates
- Adapt product to calibration product
39
Time constant 1
< 1s
Time constant 1 too small
- Input value has to be > 1
40
Time constant 2
< 1s
Time constant 2 too small
- Input value has to be > 1
41
min. measuring
range not reached
Enter negative measuring range
- Check final value
(only with square or cubic curve)
42
max. measuring
range exceeded
Measuring range input lies above
the limit
- Check final value
(only with square or cubic curve)
43
Rate Overflow
With twisted measuring range
- Check input
44
Current input error Input current above 23 mA
- Check power source
46
Check external
mass counter set-
ting
Input is outside the limits and is
ignored.
(triggered when changing parame-
ters – measuring range or mass
counter)
- Increase step-width of mass counter or
reduce maximum measuring range
(safety circuit, so relay output can still
follow)
47
Check external
mass counter set-
ting
Input is outside the limits and is
ignored.
(triggered when changing parame-
ters – measuring range or mass
counter)
- Increase step-width of mass counter or
reduce maximum measuring range
(safety circuit, so relay output can still
follow)
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...