6.2 Software Functions and System Configuration
66
6.2.1 Slave Menu Functions
General Data
Password:
You can enter a number comprising up to 6 digits. Upon confirma-
tion with <
enter
>, the system is protected against unauthorized
manipulation of the parameters. All parameters can still be viewed,
however. With password protection enabled, the <
run
> button is
locked. External measurements such as batch and tare measure-
ments can be started via the digital inputs. Password protection is
canceled when you enter the correct numerical value and confirm
your entry with <
enter
>.
Date:
Enter the current date in the format TT.MM.YY. The correct date is
important for automatic correction of the activity decay of the
source.
Decay compensation is performed every night at midnight or when
pressing the <
run
> button to start the measurement modes, when
the last decay compensation took place more than 24 hours ago.
The tare count rate and the count rates of the individual calibration
points are corrected.
Time:
Enter the current time in the format HH.MM.
System/Version:
Display of instrument type and software version.
Detector software version:
Display of the software version of the connected detector.
Language:
You can choose German, French, Spanish and English as dialog
language by pressing the <
^^^
> key
Print Parameter:
A printer with serial interface can be connected on the front panel
to document all parameter settings. The printout in the appendix
shows an example. A PC may also be connected and the data
transferred to the computer and processed there using a terminal
program.
See Example printout on page 128.
Factory Setting:
All parameters can be reset to the factory setting by pressing
<
sk1
> and <
clear
> at the same time.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...