Section 3 Installation
27
B) Mounting the Detector
Co or Cs measuring frames
□
Dismantle the mounting set enclosed with the instru-
ment.
□
Fix the detector to the holding plate using the clamps
and the hexagon head cap screws.
□
Fix holding plate using the two fastening bows in the
center of the top part of the measuring frame. The win-
dow of the detector head must face the source and be
located vertically above the source. The window is ad-
justed correctly if it is in the dead center of the measur-
ing frame.
Americium measuring frame
□
The detector is fixed to the Americium measuring frame
using the clamping device attached to the centerpiece.
□
The detector is installed properly if it extends about
75 mm beyond the frame measured from the lower edge
of the centerpiece.
□
Readjust the detector after you have connected and
turned on the evaluation unit and opened the radiation
exit channel: The highest count rate is displayed in the
correct position (see section Re-adjusting the Detector).
3.2.2 Measuring Frames on Screw Conveyors
The same instructions apply for the installation of measuring frame
and detector on a screw conveyor.
The measuring frame should be installed in a location where the
material in the screw conveyor flows smoothly. This means that
the conveyor material in the screw conveyor must not fall back be-
hind the screw shaft, as otherwise the material speed cannot be
measured accurately. The material flow can be optimized by in-
creasing the conveyor throughput speed.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...