6.2 Software Functions and System Configuration
64
Moisture %
Moisture reading when the radiometric bulk flow measurement is
configured accordingly and a moisture meter is connected (only in
operating mode
with moisture compensation
).
Dry totalizer (t or kg)
Display of dry weight in the selected unit (only in operating mode
with moisture compensation
).
Raw data
By pushing the <
more
> button, the raw data of the connected de-
tectors can be displayed live. The following raw data can be read
for each detector:
HV
Shows the actual high voltage value at the detector.
HV-auto: Automatic HV control is enabled. This is the normal mode
for the HV.
HV test: A manual fixed value has been set in the Service menu.
This setting should only be used for a test.
cps
Display of the unaveraged live count rates of this detector.
Det. Temp.
Display of the detector temperature.
Service Menu
See Service Functions on page 95.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...