Section 12 Radiation Protection
113
12.3.1 Exposure when Installing the Shielding
Provided that any work in the vicinity of the shielding is prepared
well, it will not take more than 20 minutes. With an assumed dose
rate of 30 microSv/h, the calculated dose is
10 µSv
.
If you compare this dose with the permissible annual doses of
5 mSv for operating staff, this work can be carried out
100 times
per year
by one and the same person.
12.3.2 Exposure when Replacing Rod Sources
It is important to calculate the possible radiation exposure even
before mounting or dismantling rod sources. An exact calculation is
possible using the equation above.
The anticipated working hours should be spilt up in work in the di-
rect vicinity of the shielding during mounting and dismantling the
source holders and work with the unshielded source while fixing
and dismantling the source and the Ms extension bar.
Again, a rather simplified estimation is possible when the work is
prepared well. Based on the assumptions of a mean distance of
0.5 m for the whole body radiation and the time you are working
with the unshielded source of 6 minutes (= 1/10 hour), the radia-
tion exposure can be calculated for different activities (A) as fol-
lows:
Dose D = A * 0.15 at Co-60
For multi-part rod sources, the estimated radiation exposure has to
be multiplied with the number of source parts.
Calculation example
A single part rod source with an activity of 400 MBq (approx.
11 mCi) has to be replaced. Using the above assumptions concern-
ing distance and time and the above equation, we get the following
result:
D = 400 * 0.15 = 60
µ
Sv
The radiation exposure in the vicinity of the shielding was already
calculated to be 10
µ
Sv. The total exposure including mounting
and dismantling can then be estimated as being
70
µµµµ
Sv
for a sin-
gle part source.
Note:
If the above assumptions do no apply, the calculations have
to be corrected accordingly. Actually, it can only be another work-
ing time which has a proportional effect on the result of the calcu-
lated dose rate.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...