12.4 Emergency Instructions
114
12.4 Emergency Instructions
In case of serious operational trouble, such as fire or explo-
sion, which could adversely affect the radiometric measur-
ing facility, it cannot be ruled out that the function of the
shielding lock, the shielding efficiency or the stability of the
source capsule have been impaired. In this case, it is possi-
ble that persons in the vicinity of the shielding have been
exposed to higher levels of radiation.
If you suspect such a severe malfunction, the Radiation
Safety Officer has to be notified immediately.
He will then in-
vestigate the situation immediately and take all necessary provi-
sions to prevent further damage and to avoid more exposure of the
operating staff to radiation.
The Radiation Safety Officer has to make sure that the measuring
system is no longer in operation and then take appropriate steps.
He may have to inform the authorities or contact the manufacturer
or supplier of the measuring system.
If adequate know-how as well as suitable instruments are avail-
able, emergency measures may be taken immediately. In this
case, proceed as follows:
1. Locate the shielding.
2. Check the function of the shielding.
3. Check the efficiency of the shielding by measuring the dose
rate.
4. Secure and label radiation protection areas.
5. Secure the shielding with source.
6. Document the event and estimate the possible radiation level
to which the persons involved were exposed.
If you suspect any damage to the source capsule, the following
points have to be observed as well:
1. Avoid contamination.
2. Take hold of source using a tool (a pair of pliers or a pair of
tweezers) and put both (source and tool) into a plastic bag.
3. Secure them behind an auxiliary shielding (concrete wall, steel
or lead plate).
4. Check if the environment is free of contamination.
5. Make sure the radioactive waste is secured and disposed off in
compliance with the pertinent regulatory requirements.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...