7.1 Background
68
Section 7 Calibration
7.1 Background
If you are working with rod detectors, the background, i.e. the
natural environmental radiation, has to be determined. The count
rate of the background has to be regarded as constant and may
therefore not be compensated for automatically. The background
should be measured by placing the detector in a certain distance
from the measuring frame (with Cs137 3 m, with Co60 at least
5 m). If the detector cable is not long enough, a two-wire exten-
sion cable may be connected. Before measuring the zero count
rate, the detector has to be in operation or operating voltage has
to be applied for at least 30 minutes.
See also Background for rod detector on page 59.
Note: If the background is not measured correctly, a display devia-
tion may occur in the course of time.
7.2 Tare Measurement
7.2.1 General Information
Before taking the measuring system into operation, the zero or
tare weight has to be determined.
The tare weight changes constantly as a result of conveyor irregu-
larities, such as variation of the thickness of belt or, as in the case
of chain conveyors, differences between chains links and scrapers.
During tare determination, therefore, the conveyor must run and
the average value must be determined over several complete con-
veyor belt cycles. The rule is: the more cycles, the more precise
the result.
The tare measurement can be checked by running a test meas-
urement with empty belt. During the test measurement (approx. 5
minutes) no or only a negligible mass (t) should be integrated.
There are two ways of performing a tare measurement:
a) Tare measurement via time
b) Tare measurement via belt length
a) Tare Measurement via Time
First, the time for one cycle has to be determined by marking the
belt and taking the time until the mark reappears. With most con-
veyors, averaging over one belt length is insufficient. For this rea-
son, depending on the available time, enter a multiple of this
value.
Example: 1 cycle measured = 56 s
15 cycles = 840 s
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...