Section 6 Evaluation Unit LB 442
59
Product Select
No. 1 to 10
Select the calibration product and confirm with
<enter>
. Please
note that the product number set in the
Measuring Parameter
menu is identical with the product number selected here. Other-
wise, no calibration measurement is possible, but only manual en-
try of the calibration data sets since the <
run
> button is locked.
Data Input
Background for rod detector
If a rod detector is used, you have to enter the background. The
count rate is read in over a period of approx. 60 s. The measure-
ment is started and stopped by pushing the <
run
> button. The
value confirmed with <
enter
> is saved.
For more information on this subject please see Background on
page 68.
1. Value to 10. Value
For calibration of the radiometric bulk flow measurement you can
enter or read in max. 10 calibration points. You need at least 1
calibration point, but you have to run through all 10 points.
Calibration points with zero values are not calculated. If a calibra-
tion point includes a count rate with „0“ cps, this and all following
calibration points are not used for calculation.
Values confirmed with <
enter
> are stored, so that the individual
parameters and measured values are always available. Therefore,
the calibration points can be entered at different times.
With the <
enter
> button you can scroll through the individual
units of measure for the respective calibration points. The cursor
position indicates the active row.
Press the <
more
> button to display the next calibration point.
Tip: Write down all values of the calibration points immediately.
In this manner, you can easily correct incorrect entries in the cali-
bration points.
Before recording a calibration point, you should determine
and enter the tare count rate. In the other case, also when
the tare count rate is entered later, the count rates in the
calibration points change.
See also chapter 7.2.2 Tare Compensation on page 70.
Summary of Contents for LB 442
Page 1: ...Radiometric Bulk Flow Meter LB 442 Id No 32815BA2 Rev Nr 02 10 10 2001...
Page 2: ......
Page 118: ...13 1 Certificates 116...
Page 119: ...Section 13 Appendix 117...
Page 120: ...13 1 Certificates 118...
Page 121: ...Section 13 Appendix 119...
Page 122: ...13 1 Certificates 120 13 1 2 ATEX Certificate for Detector...
Page 123: ...Section 13 Appendix 121...
Page 124: ...13 1 Certificates 122...
Page 125: ...Section 13 Appendix 123...
Page 126: ...13 1 Certificates 124 13 1 3 ATEX Certificate for Wall Housing...
Page 127: ...Section 13 Appendix 125...
Page 128: ...13 1 Certificates 126...
Page 129: ...Section 13 Appendix 127 13 1 4 Declaration of Conformance...
Page 132: ...13 3 Tachometer Installation 130 13 3 Tachometer Installation...
Page 133: ...Section 13 Appendix 131...
Page 134: ...13 3 Tachometer Installation 132 13 3 1 Tachometer...
Page 138: ...13 4 Mechanical Drawings 136 13 4 4 Free Fall Measurement...
Page 139: ...Section 13 Appendix 137 13 4 5 Point Detector...
Page 145: ...Section 13 Appendix 143 13 4 11 Evaluation Unit...