4.10 Option I/O errors
Table 27: Option I/O errors description
Error Code
Error Message
Description
A1:AnalogCh(S2:3) Error!
ADC Channel(S2:3) is not responding
Analog input /RTD input is not working.
If error persists after power cycle,
contact Panametrics factory
A2:AnalogCh (S2:4) Error!
ADC Channel(S2:4) is not responding
Analog input /RTD is not working. If error
persists after power cycle, contact
Panametrics factory
A3:AnalogCh (S2:1) Error!
DAQ Channel (S2:1) is not responding
Analog output (4-20mA) is not working.
If error persists after power cycle,
contact Panametrics factory
A4:AnalogCh (S2:2) Error!
DAQ Channel (S2:2) is not responding
Analog output (4-20mA) is not working.
If error persists after power cycle,
contact Panametrics factory
A6:(S2:3)Ch Not Calibrated
Error occurs when Analog Input/
RTD(S2:3) are not calibrated
Calibrate the Analog Input/RTD input. If
error persists after calibration, contact
Panametrics factory
A7:(S2:4)Ch Not Calibrated
Error occurs when Analog Input/RTD
(S2:4) are not calibrated
Calibrate the Analog Input/RTD input. If
error persists after calibration, contact
Panametrics factory
A10:(S2:3)Input NotConnect!
Analog Input: Error occurs when
(4-20mA) input is not connected at
Channel (S2:3).
RTD Input: Error occurs when RTD input
is not connected or temp greater than
390 deg C at Channel (S2:3)
Check connectivity for Analog Input/
RTD input and RTD temperature.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
A11:(S2:4)Input NotConnect!
Analog Input: Error occurs when
(4-20mA) input is not connected at
Channel (S2:4).
RTD Input: Error occurs when RTD input
is not connected or temp greater than
390 deg C at Channel (S2:4)
Check connectivity for Analog Input/
RTD input and RTD temperature.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
A12:(S2:3)Ch OverRange Err!
Exceeds input values. For analog input
(S2:3) greater than 21mA
Ensure analog input current less than
21mA. Try power cycling the meter.
If error persists after power cycle,
contact Panametrics factory
A13:(S2:4)Ch OverRange Err!
Analog input(S2:4) greater than 21mA
Ensure analog input current less than
21mA. Try power cycling the meter.
If error persists after power cycle,
contact Panametrics factory
64
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...