Table 17: Lining Materials
Lining Material
Other
Tar Epoxy
Glass Pyrex
Asbestos Cement
Mortar
Rubber
Teflon
3.6.2 Programming the Fluid
The Fluid menu (see Figure 69) allows the user to specify
all the parameters of the fluid flowing through the pipe
that is required to ensure accurate ultrasonic flow rate
measurements. Use steps as in section “Log-in and Primary
Pages” to navigate to the Programming page.
1. Highlight [Composite] and press [ENTER]. Scroll down,
highlight [Fluid] and press [ENTER].
2. Highlight [Density], press [ENTER] and program the
actual density [Density (Act)] and reference density
[Density (Ref)] of the process fluid.
3. Then highlight [Kinematic Viscosity], press [ENTER] and
program the kinematic viscosity of the process fluid.
4. Then highlight the [Tracking] option. The tracking
window is used to scan through the speed of sound
range programmed to detect the signal when the
user is unsure of the fluid sound speed. The meter also
supports a list of standard fluid types. If the process fluid
is not listed in standard fluids list and if you are unsure
of the fluid sound speed, set tracking window to On and
program the minimum and maximum sound speed
range to scan.
5. For standard fluids list (see Table 18) supported in the
meter, the minimum, maximum and nominal sound
speeds are automatically selected.
6. Also program the process [Fluid Temperature] and
[Ambient Temperature] see Advanced programming
section.
Table 18: Standard Fluid List
Tracking On
Tracking Off
Other
Other
Water (0 to 260 C)
Water (0 to 260C)
LNG
LNG
Oil 22C
Oil 22 C
Sea Water
Lube Oil
Crude Oil
Methanol (20 C)
Table 18: Standard Fluid List
Tracking On
Tracking Off
Ethanol
Freon R12
Diesel
Gasoline
Liquid Nitrogen (-199C)
45
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...