H.11 Error handling
The flow meter publishes the error status on the Fieldbus
along with the real data. The error status can be seen in the
CH_x_Reported Error parameter on the Channel Transducer
Block. In addition, the Quality parameter shown with each of
the process variables reports the error. In Figure 139 below,
the CH1_REPORTED_ERROR shows as E1.
Note:
For more information on the actual measurement
errors and possible causes, see Chapter 4, Error Codes and
Troubleshooting.
Figure 139: Reported error
Whenever the meter is in measurement error, the quality
bit for the published parameter shows bad quality (see
Figure 140 below). To change the quality bit to good, the
measurement error at the meter must be removed.
Notice the
QUALITY.STATUS
field shows as
Bad
and the
SUBSTATUS
field shows as
Sensor Failure
. This information
indicates a measurement error which must be corrected.
Figure 140: Quality bit error
H.12 Simulation mode
Simulation
mode allows the user to test the FF
implementation without the instrument providing real data.
The meter PCB is shipped with simulation mode disabled. To
enable simulation mode, complete the following steps:
CAUTION!
To prevent damage to the electronic
components, always use ESD protection
whenever handling printed circuit boards.
1. Remove the PCB from the meter.
2. Locate the jumper P5 (see Figure 141 below).
3. Move the P5 jumper to the left one place (pins 2 and 3)
to enable simulation mode.
4. Re-install the PCB into meter
5. Verify that the Block error field in the Resource Block
shows SimulationActive status.
Figure 141: P5 on XMT1000 PCB
123
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...