Figure 69: Fluid programming
3.6.3 Programming the Path Configuration
Use steps as in section “Log-in and Primary Pages” to navigate to the Programming page. Refer to Figure 70 for the Path
configuration options.
1. Highlight [Composite] and press [ENTER]. Scroll down and select [Path] and press [ENTER].
2. Select [Path Configuration], [Path Weights] and [Path Error Handling].
3. Path Weights are used in Composite flow velocity calculations as in the following equation:
4. If [Path Error Handling] is set to On, the meter will continue to provide measurements even if one or two channels are in
error. Unless, all three channels (for 3 Path (TD-TD-TD)) are in error the flow measurement continues.
Figure 70: Path configuration
46