Table 24: System error description and recommended actions
Error Code
Error Message
Description / Recommended Action
S1: In Config Mode
In configuration mode indicator
Indicator:
This is displayed when a user
has logged in to either Operator, Admin
or Factory access level. The indicator
will clear automatically when the user
logs out or saves the configuration
changes
S2: Invalid User
Invalid user warning
Warning:
The passcode entered for
access level is incorrect. Please log
in with the correct access level and
passcode
S3: Invalid Request
Invalid request warning
Warning:
An invalid communication
packet was received and discarded.
Or, the requested operation is invalid.
Please send a valid packet or operation
request
S4: Invalid Param Range
Invalid parameter range warning
Warning:
The value programmed for
the parameter was out of range and
hence discarded. Please enter a valid
range
S5: Unsupported Parameter
This parameter is not supported
Warning:
A read or write request to an
unsupported parameter was received
S6: Flow Measurement
One or more flow measurement
channels are in error
Error:
One or more flow measurement
channels are in error; accuracy of the
measurement may be compromised.
For more details please check flow(E)
errors
S7: Persistent Param CRC
Persistent parameter CRC fault
Fault:
Persistent parameter CRC failed.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
S11: Clock Frequency
Clock frequency error
Fault:
Input clock frequency failure.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
S12: CPU
CPU error
Fault:
CPU registers have stuck bits.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
S13: Invariable Flash Memory
Flash memory fault
Fault:
Flash memory test failed. Try
power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
60
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...