4.3.1b Multi-channel error
If more than one channel is in error, the most likely cause is changes in process/flow conditions such as:
1. Excessive turbulence.
2. Discontinuities in fluid characteristics such as multi-phase flow, flashing, pockets of gas, presence of bubbles or solid
particles, cavitation or rapidly changing fluid type.
3. Extreme fluid properties, such as pressure or temperature.
4. Wax build-up inside the pipe.
5. Partially filled pipe.
After you have tried eliminating/correcting for any most likely causes mentioned above, if error still exists, also check:
1. Incorrect programming on Error Limits or flow condition changes that now make previous programming invalid.
2. Defective/Damaged cables, transducers, incorrect physical spacing, couplant, buffer or electronics.
In case you are unable to clear the errors, collect diagnostic data and parameter information for each channel in the
Diagnostics Data Table prior to calling your local sales or service center.
4.3.1c Viewing channel specific error/warnings
To indicate the health of the meter, PanaFlow™ LC has built-in Error codes. The Channel specific errors are very critical
in determining the corrective actions required. Figure 80 below shows the steps to view current channel specific errors/
warnings. The description of the Error Codes and the recommended actions are provided in Table 21 below.
Main Menu
Errors/Warning
1/1
Errors/Warning
Flow Info
Composite Flow Info
E23: MultiChAccuracy
Velocity
0.000
m/s
Display Format
Program
Flow Info
Ch 1 Flow Info
Ch 2 Flow Info
Ch 3 Flow Info
CH1 E1: SNR Error!
CH2 E1: SNR Error!
CH3 E1: SNR Error!
Figure 80: Viewing current channel specific errors
54
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...