Table 28: Diagnostic Parameter Description and Health Indicators
Parameter
Description
Good
Bad
Lining Time
Transit time inside the pipe
lining
N.A
If the value is negative, then there
is an indication of problem with the
configuration parametersw
Signal Quality
Up
Signal quality of the
upstream transducer
>1000
<1000
Signal Quality
Down
Signal quality of the
downstream transducer
>1000
<1000
Amplitude Up
Signal amplitude of the
upstream transducer
>14 and <32
>32 or <14
Amplitude Down Signal amplitude of the
downstream transducer
>14 and <32
>32 or <14