3.5.3d Setting up alarm output
To program an Alarm Output, use steps as in section “Log-in and Primary Pages” to navigate to the Input/Output settings
page:
1. In Meter Menu highlight [Digital Output (Std)] and press [ENTER]. Then select [Alarm] option.
2. Select the [Alarm State], [Alarm type], [Measurement] and [Alarm Value] options to suit your needs. The Figure 60 below,
shows the available options. Also refer to the Table 13 below to understand each option.
For measurement options available on Alarm Output refer to Table 12.
Table 12: Measurement options for Alarm output
Measurement Channel
Measurement Options for Frequency Output
Composite
Velocity, Actual Volumetric, Standard Volumetric, Mass Flow,
Average Volumetric Flow Rate, Sound speed, Reynolds #
Channel x
Velocity, Actual Volumetric, Standard Volumetric, Mass
Flow, Average Volumetric Flow rate, Sound speed, Standard
Deviation of Gain, Gain and SNR
Table 13: Alarm Output Options
Option
Functional Description
Alarm State
Choose if the Alarm State should be Normally Open,
Normally Closed or Failsafe
Alarm Type
For Alarm State selected as Normally Open or Normally
Closed, Alarm Type can be set to High or Low. If set to high,
Alarm will be triggered if the selected Measurement goes
above the programmed Alarm Value
Measurement
Choose the channel and Measurement that is monitored for
Alarm Trigger
Alarm Value
Enter measurement value that should be a trigger point
36
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...