4. Use the pressure bolt to secure the transducer in place. The pressure bolt should fit into the dimple on the transducer.
Hand tighten the bolt just enough to hold the transducer in place, but do not overtighten it or the fixture will lift off the
pipe.
5. Repeat Steps 1-4 to mount the other transducer in the fixed short block. See Figure 22 below for completed typical
clamping fixture installations.
6. Connect the other end of the transducer cables to the junction box in Section 4: Transducer Wiring of this manual.
Note:
If the transducers are mounted into the clamping fixtures properly, the two transducer cable connectors will face away
from each other as shown above.
Figure 22: SCF clamping fixture (top left), UCF clamping fixture (top right), GCF clamping fixture (bottom left), MCF clamping fixture (bottom right)
11
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...