3.5.3b Setting up frequency output
To program a Frequency Output, use steps as in section
“Log-in and Primary Pages” to navigate to the Input/Output
settings page:
1. In Meter Menu highlight [Digital Output (Std)] and press
[ENTER]. Then highlight [Frequency] option and press
[ENTER].
2. Setup the [Measurement], [Base Value], [Full Value], [Full
Frequency] and [Error Handling] options to suit your
needs. The Figure 59 below, shows the available options.
Also refer to the Table 9 below to understand each
option.
For measurement options available on Frequency Output
refer to Table 8.
Table 8: Measurement options for frequency output
Measurement
Channel
Measurement Options for Frequency
Output
Composite
Velocity, Actual Volumetric, Standard
Volumetric, Mass Flow, Average
Volumetric Flow Rate, Sound speed,
Reynolds #
Channel x
Velocity, Actual Volumetric, Standard
Volumetric, Mass Flow, Average
Volumetric Flow rate, Sound speed,
Standard Deviation of Gain, Gain and SNR
Table 9: Frequency output options
Option
Functional Description
Measurement Choose the channel and Measurement
to output
Base Value
Enter measurement value that should
correspond to the min. value of the
frequency range
Full Value
Enter measurement value that should
correspond to the max. value of the
frequency range
Fspan
Enter the highest value of the Frequency
range you want to output
Error Handling
Choose the Frequency output response
during an error condition: Low, High, Hold,
or Other Value. During a meter error, the
chosen Error Handling value will be sent
to the Frequency Output.
33
Summary of Contents for Panametrics PanaFlow LC
Page 1: ...PanaFlow LC User s manual 910 327 Rev A...
Page 2: ...ii...
Page 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Page 4: ...no content intended for this page iv...
Page 12: ...no content intended for this page xii...
Page 44: ...Figure 58 Setting up pulse output 32...
Page 49: ...Figure 60 Setting up alarm output 37...
Page 80: ...no content intended for this page 68...
Page 82: ...no content intended for this page 70...
Page 86: ...no content intended for this page 74...
Page 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Page 104: ...no content intended for this page 92...
Page 106: ...no content intended for this page 94...
Page 114: ...no content intended for this page 102...
Page 138: ...no content intended for this page 126...
Page 140: ...no content intended for this page 128...