29
5-2. Three-month Inspection
* This section includes references to the oxygen controller and the weight monitor. If your unit is not
equipped with these functions, skip over such references.
Check the following points every three months or so.
Item to check
Procedure
Description
Incubator Air
Temperature
Control
Set the temperature to 36.0
℃
in manual
control.
The displayed incubator air
temperature should be stable at
36.0±1
℃
.
Skin
Temperature
Control
Place the skin temperature probe at about
10cm above the center of the mattress surface
in the incubator and set the temperature to
36.0
℃
in servo control.
The displayed skin temperature
should be stable at 36.0±0.5
℃
.
Humidity Control
Set RH to 90% with the incubator air
temperature set to 32.0
℃
.
The displayed relative humidity
should be stable at 90
±
5%.
Oxygen
Concentration
Control
(1) Supply oxygen at 10L/min from the oxygen
connecting port 1.
(2) (The unit equipped with the oxygen
controller)
Connect oxygen to the oxygen connecting
port 2 and, after 21% calibration, set the
oxygen concentration to 40%.
(1)The oxygen concentration in
the incubator should rise to
65% or higher.
(2)It should be possible to carry
out 21% calibration
accurately, and the
displayed oxygen
concentration should be
stable at 40±2%.
Pulse oximeter
Connect a Masimo Sensor to the sensor
module, and attach the sensor to the inspector.
The displayed SpO
2
should be
95~100% and the pulse rate
should also be displayed.
Fan
Check the exterior appearance by visual
inspection.
The fan should be attached
correctly and intact without any
breakage or deformation.
Humidifier
(1) Install a cartridge tank without any water
in it.
(2) Open the humidity chamber cover.
(3) Draw out the humidity chamber by about
3cm.
(4) Fit in the humidity chamber with the boiler
cap off.
(1) Check that a low water level
alarm appears.
(2) ~(4) Check that a humidity
chamber off alarm appears.
Power Failure
Alarm
Turn the power switch on, and then remove
the power plug from the power outlet.
An audible alarm should sound
and the power failure alarm
indicator should come on.
Weight Monitor
(The unit equipped with the weight monitor)
Put a 5kg weight on the center of the mattress,
measure the weight, and check the displayed
value.
The displayed value should be
in the range of 5000±5g (when a
5kg weight is used).
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...