168
Division
Product inspection
Serial No.
Section
Oxygen monitor
Tools and jigs to be used
No Inspection
procedure
Remarks
3
Oxygen flow rate alarm
Connect the piping connecting hose to the oxygen connecting port 2.
Press the O
2
swtich, and set the oxygen concentration to 40%. (At this time,
do not supply oxygen.)
Check that the oxygen flow rate alarm message appears and that an audible
alarm sounds.
* After confirming the above, turn the oxygen function off, and then turn it
on again to set the incubator in the oxygen monitor mode.
4
Oxygen sensor alarm
Remove one oxygen sensor from the sensor module.
Check that the oxygen sensor alarm message appears and that an audible
alarm sounds.
Put the removed oxygen sensor back, and perform 21% calibration.
Do the same on the other oxygen sensor.
Oxygen flow rate alarm
Oxygen sensor alarm