172
Division
Product inspection
Serial No.
Section
Temperature
Tools and jigs to be used
No Inspection
procedure
Remarks
3
Skin Temperature 1
Connect the skin temperature probe to the Skin Temperature 1, and place
the thermistor on the mattress in the hood.
Press the servo control switch to change to the servo control mode.
Set the skin temperature to 36.0
℃
.
After 10~15 minutes, check that the reading on the skin temperature 1
display is stable at 36.0
±
0.1
℃
.
Press the Screen Changeover switch to have the Trend 1 screen displayed.
Check that the plot on the skin temperature 1 graph in the upper portion of
the screen is the sam as the reading on the skin temperature 1 display.
Check that the reading on the skin temperature 1 display on the right top of
the screen is the same as the skin temperature 1 shown on the numerical
value screen.
Check that the graph is recorded.