13
Note :
⑬
and
⑭
are available only on the unit with
oxygen controller.
⑮
is available only on the unit with built-in pulse
oximeter.
⑯
,
⑰
and
⑱
are available only on the unit with
weight monitor.
Name
⑫
Relative Humidity Setting Switch
Press this switch to set a relative humidity.
When this switch is pressed, the selection of
humidity message will be displayed. and a
relative humidity can be set.
Select a desired setting by pressing an
appropriate setting switch.
⑬
ON/OFF Switch
Press this switch to turn the oxygen
controller on/off.
⑭
O
2
Switch
Press this switch to set an oxygen
concentration. When this switch is pressed,
the selection of oxygen concentration
message will be displayed, and an oxygen
concentration can be set.
Select a desired setting by pressing an
appropriate setting switch.
⑮
Alarm Limit Selector Switch
Press this switch to set an SpO
2
/pulse rate
alarm limit.
Every time this switch is pressed, selection
of a SpO
2
alarm message and the selection
of pulse rate alarm massage will be
displayed by turns, and an alarm limit for
the mode currently being displayed can be
set.
Select a desired setting by pressing an
appropriate setting switch.
⑯
Weight Switch
Press this switch to weigh or store the
measurement value.
⑰
Tare Switch
Press this switch to tare.
⑱
Mode Switch
Press this switch to change the weight
measurement mode.
⑲
Display
Displays an air temperature, a relative humidity
and an oxygen concentration in the incubator,
skin temperature, %SpO
2
, pulse rate and weight
of the infant etc. on the Numerical value screen,
Trend 1 screen and Trend 2 screen.
Name
①
Alarm Indicator
Flashes or illuminates when an alarm
condition occurs.
②
Power Failure Alarm Indicator
Illuminates when power supply is
interrupted due to power breakdown or a
disconnected power plug etc.
③
System Alarm Indicator
Illuminates when any system error is
detected by the self-diagnosis.
④
Screen Changeover Switch
Every time this switch is pressed, the
screen is changed.
⑤
Brightness Adjusting Switch
Press this switch to adjust the brightness
of the screen.
⑥
°C/°F Selector Switch
Press this switch to change the temperature
display unit from °C to °F and vice versa.
⑦
Setting Switch
Press this switch to set temperature,
relative humidity, SpO
2
/pulse rate alarm
limits, etc. to a desired level.
Every time
switch is pressed, a
setting increases.
Every time
switch is pressed, a
setting decreases.
⑧
Alarm Silence/ reset Switch
Press this switch to silence an audible
alarm or reset an alarm condition.
⑨
Manual Control Switch
Press this switch to operate the incubator
in the Manual Control mode.
⑩
Servo Control Switch
Press this switch to operate the incubator
in the Servo Control mode.
⑪
Override Switch
Press this switch while the Servo
indicator is flashing to select the override
mode; the
“
>37.5°C
”
indicator will flash,
and the high skin temperature range of
37.6-39.0°C will be available.
Press this switch while the Manual
indicator is flashing to select the override
mode; the
“
>37.0°C” indicator will flash,
and the high incubator air temperature
range of 37.1-39.0°C will be available.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...