27
(Calibration)
Before calibration, obtain a 5kg reference weight.
(1) Check that the power switch of the incubator is off.
While pressing the calibration switch on the right
side of the main body with a pointed object, turn on
the power switch.
(2) The message “INITIATING CALIBRATION” will
appear in the message area at the bottom of the
display, and the weight monitor will proceed to the
calibration mode.
(3) The 0kg calibration procedure will start
automatically. Be sure to leave nothing on the
mattress platform. Wait for the numerical value on
the display to stabilize in 5~10 seconds, and then
press the Mode switch. The 0kg calibration
procedure will end, and the weight monitor will
proceed to the 5kg calibration mode. If it doesn’t,
see the step (6) below.
(4) The message “CALIBRATING TO 5kg” will appear
in the message area. Then put the 5kg reference
weight in the center of the mattress platform. Wait
for the numerical value on the display to stabilize
in 5~10 seconds, and then press the Mode Switch.
The 5kg calibration procedure will start.
(5) When the 5kg calibration procedure is completed, a
message “Calibration completed” will appear and
the calibrated value will be shown on the weight
display. The value displayed right after the
calibration may differ from the calibrated value by
several grams. After confirming the message and
the weight reading, press the Mode Switch. The
message “END OF CALIBRATION MODE” will
appear, and the weight monitor will proceed to the
manual mode.
Calibration switch
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...