132
Division
Assembling the main body
Serial No.
Section
Attaching the filter case and the oxygen
controller/the oxygen supply port
Descriptions
Tools and jigs to be used
(1) Phillips screwdriver
No Part
name Qty
No
Assembling procedure
Remarks
1
Main body case 2200
1
1
Fix
②
to
①
with the screws of
②
.
Check that the electrostatic filter
has been attached when fixing.
2
Filter case
1
3
Oxygen controller/oxygen supply port
1
2
Fit
③
in
①
. Fix
⑤
to
③
with
⑥
.
(1)
4
Small truss screw (M4×8)
2
3
Connect
⑦
to the joints of
③
and
the oxygen supply valve of
①
.
After connecting, check that
⑦
will not come off.
5
Ground cord 63
1
6
Sems (M3×8)
1
4
Fix
③
to
①
with
④
.
(1)
7
Hydraulic tube
1
5
8
6
9
7
10
8
※③
applies to a unit equipped with the
oxygen controller.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...