10
.
Set the temperature to 37.5
℃
in the servo
control mode.
11
.
Disconnect the skin temperature probe from
the sensor module.
Check:
(1) Does a skin temperature probe alarm
appear?
12. Disconnect the cable from J2 on the relay box.
Check:
(2) Does the system failure alarm indicator come on in less than 10 seconds?
(Checking the power failure alarm)
13. Disconnect the power cord.
Check:
(1) Does the power failure alarm indicator come on?
(2) Does an audible alarm sound?
(Preparation 2)
14. Set the power switch to the OFF position.
15. Attach the power cord.
(Weight Monitor)
16. Connect the weight monitor cord to the weight monitor input on the right side of the main body.
17. Turn the power switch on to activate the unit.
Check:
(1) Does “MANUAL WEIGHT MEASUREMENT
----
g” appear in the message area?
18. Press the Tare switch.
Check:
(2) Does the weight monitor begin to deduct the tare, and then does it enter the continuous
measurement state?
(3) In the continuous measurement state, does the numerical value change when a little force is
applied to the mattress platform?
19. Press the Weight switch.
Check:
(4) Does the weighing procedure start, and then does a message indicating the normal completion of
measurement appear?
(End of the Test)
20. Turn the power switch off.
Summary of Contents for V-2200
Page 1: ......
Page 2: ......
Page 4: ......
Page 14: ...12 4 2 Display Control Panel...
Page 187: ...185 Available only for units destined for the US...
Page 189: ...187 C60F0263 Available only for units with serial No 1680000 or higher...
Page 190: ...188 C60F1003 Available only for units destined for the US...
Page 191: ...189 9 2 HL stand...
Page 192: ...190...
Page 193: ...191 10 Replacement Parts List Replacement Parts List...
Page 194: ...192 10 1 Pole Arm Upper Part of the Main Body Accessories and Humidity Chamber Section Fig 1...
Page 196: ...194 10 2 Hood Section Fig 2...
Page 198: ...196 10 3 Main Body Fig 3...
Page 202: ...200 10 5 HL Stand Fig 5...
Page 211: ......
Page 212: ...C60S220 201109...